주요 특징

• Voltage output : 6V / 12.5V / 20V / 40V / 60V/ 100V/ 150V/ 300V/ 400V/ 600V

• Power output:1200W ~ 1560W

• C.V/ C.C priority mode

• Adjustable voltage/current rise and fall time

• 1U height and 19”Rack Mount Size

• Bleeder Control function

• Internal Resistance function

• Protection: OVP, OCP, OHP,UVL, AC Fail, FAN Fail

• Standard: USB, LAN, RS-232, RS-485, analog control

• Option: GPIB、isolated analog interface (voltage control/ current control)

 

GW Instek PSU-HV series has five models, including PSU 100-15, PSU 150-10, PSU 300-5, PSU 400-3.8, and PSU 600-2.6. The launch of PSU-HV is to complete the existing PSU series so as to satisfy high voltage application demands, allowing the augmented PSU series to cover a voltage range from 6V to 600V. PSU-HV inherits the functional design and maintains the high power density characteristic and 1U height appearance of the PSU-LV series (PSU 6-200, PSU 12.5-120,PSU 20-76, PSU 40-38和PSU 60-25). Furthermore, the original maximum output voltage of 60V is expanded to the maximum voltage of 600V and the maximum power of 1560 watts. The launch of the PSU-HV series augments the existing PSU series to fully satisfy the extensive voltage demands of 1U power supply market and provides system integrators with more flexibilities and selections to conduct system integration. The introduction of the PSU-HV series has perfected the PSU product line, which satisfies the applicationrequirements ranging from low voltage and large current to high voltage.

 

Utilizing same model units of the PSU series to conduct series and parallel connections can increase total output power, total current or total voltage. The wide voltage and current output ranges of the PSU series can fully satisfy various voltage and current measurement requirements. The PSU series is a single power output DC programmable power supply, which outputs 1200W to 1560W. The PSU series provides maximum 2 units in series connection (models under300V) to achieve maximum 600V or 4 units in parallel connection to obtain maximum800Aand the maximum output power of 6.24 kilowatts.

 

 The PSU series allows settings for CC priority or CV priority. Under CC or CV mode, users can adjust slew rate for output voltage or current based upon test requirements. There are two kinds of slew rate settings: high speed priority and slew rate priority.  High speed priority sets slew rate at the maximum speed to reach CC or CV mode. Slew rate priority allows users to set slew rate for CC or CV mode in order to control rise or fall slew rate.Slew rate priority mode is ideal for motor tests by adjusting the rise time of output voltage to protect DUT from being damaged by inrush current occurred at turn-on.

 

Comparing with other 1U power supplies available in the market, PSU supports a most complete array of interfaces, including USB, LAN, RS-232, RS-485, analog control interface, GPIB (option), isolated analog interface (voltage control), and isolated analog interface (current control). Via the multi-drop mode, PSU will not need any switch/hub and GPIB cable for remote control and slave unit augmentation when using LAN, USB or GPIB. This feature can helpusers save costs on augmentation equipment for connecting slave while using LAN or USB.

 

The new PSU-HV series is ideal for the primary input of DC/DC converter and servomotor production application. PSU is often integrated into component test systems such as aging test equipment for capacitors; 600V DC bias applications; aging test equipment for diode; semiconductor production equipment; automotive electronics; and ECU for V8 engine or V12 engine, etc.

 

The PSU series provides users with flexible settings of High/Low Level or  Trigger input /Trigger output signals with pulse width of 1 ~ 60ms. Trigger input controls PSU to output or upload preset voltage, current and memory parameters. While outputting or uploading preset voltage, current and memory parameters PSU can produce corresponding Trigger output signals.

 

Specification

PSU

6-200

12.5-120

20-76

40-38

60-25

100-15

150-10

300-5

400-3.8

600-2.6

Output ratings

Rated outputvoltage (*1) (V)

6

12.5

20

40

60

100

150

300

400

600

Rated outputcurrent (*2) (A)

200

120

76

38

25

15

10

5

3.8

2.6

Rated outputpower (W)

1200

1500

1520

1520

1500

1500

1500

1500

1520

1560

Ripple and noise(*5)

CVp-p( 10 ~20MHz) p-p (*6) (mV)

60

60

60

60

60

80

100

150

200

300

CVrms(5Hz ~1MHz) r.m.s. (*7) (mV)

8

8

8

8

8

8

10

25

40

60

CCrms(5Hz ~1MHz) r.m.s.(*12) (mA)

400

240

152

95

75

45

35

25

17

12

Load regulation

Voltage(*4)(mV)

2.6

3.25

4

6

8

12

17

32

42

62

Current(*11)(mA)

45

29

20.2

12.6

10

8

7

6

5.76

5.52

Line regulation

Voltage(*3)(mV)

2.6

3.25

4

6

8

12

17

32

42

62

Current(*3)(mA)

22

14

9.6

5.8

4.5

3.5

3

2.5

2.38

2.26

Analog Programming and Monitoring

Externalvoltage control output voltage

 Accuracy and linearity: ±0.5% of rated output voltage.

Externalvoltage control output current

 Accuracy and linearity: ±1% of rated output current.

Externalresistor control output voltage

 Accuracy and linearity: ±1% of rated output voltage.

Externalresistor control output current

 Accuracy and linearity: ±1.5% of rated output current.

Output voltagemonitor

 Accuracy: ±1%

Output currentmonitor

 Accuracy: ±1%

Shutdowncontrol

 Turns the output off with a LOW (0V to 0.5V) or short-circuit.

Output on/off control

 Possible logic selections:

 Turn the output on using aLOW (0V to 0.5V) or short-circuit, turn the output off using a HIGH (4.5V to5V) or open-circuit.

 Turn the output on using aHIGH (4.5V to 5V) or open-circuit, turn the output off using a LOW (0V to

Alarm clearcontrol

 Clear alarms with a LOW (0V to 0.5V) or short-circuit.

CV/CC/ALM/PWRON/OUT ON indicator

 Photocoupler open collector output;Maximum voltage 30V, maximum sink current 8mA.

Trigger out

 Maximum low level output = 0.8V; minimumhigh level output = 2V; Maximum source current = 8mA.

Trigger in

 Maximum low level input voltage = 0.8V;minimum high level input votage = 2V, Maximum sink current = 8mA.

Front Panel

Display, 4digits,
Voltage accuracy (0.1%+mV)

12

25

40

80

120

200

300

600

800

1200

Current accuracy (0.2%+mA)

600

360

228

114

75

45

30

15

11.4

7.8

Indications

 GREEN LED's: CV, CC, V, A, VSR, ISR, DLY, RMT, LAN, M1, M2, M3,RUN,OutputON

 RED LED's: ALM, ERR

Buttons

 Lock/Local(Unlock), PROT(ALM_CLR),Function(M1), Test(M2), Set(M3), Shift, Output

Knobs

 Voltage, Current

USB port

 Type A USB connector

Transient response time (*10)

Transientresponse time (ms)

1.5

1

1

1

1

1

2

2

2

2

Output response time

Rise time (*8)

 Rated load(ms)

80

80

80

80

80

150

150

150

200

250

 No load (ms)

80

80

80

80

80

150

150

150

200

250

Fall time(*9)

 Rated load (ms)

10

50

50

80

80

150

150

150

200

250

 No load (ms)

500

700

800

1000

1100

1500

2000

2500

3000

4000

Programming and Measurements (RS-232/485, USB, LAN, GPIB)

Programming accuracy

 Outputvoltage  (0.05%+mV)

3

6.25

10

20

30

50

75

150

200

300

 Output current(0.2%+mA)

200

120

76

38

25

15

10

5

3.8

2.6

Programming resolution

 Output voltage(mV)

0.2

0.4

0.7

1.3

2

3.4

5.2

10.2

13.6

20.4

 Output current(mA)

6

4

2.5

1.2

0.8

0.5

0.34

0.19

0.13

0.09

Measurement accuracy

 Output voltage(0.1%+mV)

6

12.5

20

40

60

100

150

300

400

600

 Output current(0.2%+mA)

400

240

152

76

50

30

20

10

7.6

5.2

Measurement resolution

 Output voltage(mV)

0.2

0.4

0.7

1.3

2

3.4

5.2

10.2

13.6

20.4

 Output current(mA)

6

4

2.5

1.2

0.8

0.5

0.34

0.19

0.13

0.09

Temperature coefficience

Voltage (ppm/℃)

 100ppm/℃ after a 30 minute warm-up

Current (ppm/℃)

 100ppm/℃ after a 30 minute warm-up

Remote sense compensation voltage(single wire)

voltage

1

1

1

2

3

5

5

5

5

5

Protection Function

Over voltage protection(OVP)

 Setting range(V)

0.6 - 6.6

1.25 - 13.75

02월 22일

4 - 44

5 - 66

5 - 110

5 - 165

5 - 330

5 - 440

5 - 660

 Settingaccuracy (mV)

60

125

200

400

600

1000

1500

3000

4000

6000

Over current protection(OCP)

 Setting range(A)

5 - 220

5 - 132

5 - 83.6

3.8 - 41.8

2.5 - 27.5

1.5 - 16.5

1 - 11

0.5 - 5.5

0.38 - 4.18

0.26 - 2.86

 Settingaccuracy (mA)

4000

2400

1520

760

500

300

200

100

76

52

Under voltage limit(UVL)

 Setting range

0 - 6.3

0 - 13.12

0 - 21

0 - 42

0 - 63

0 - 105

0 - 157.5

0 - 315

0 - 420

0 - 630

Overtemperature protection(OHP)

 Turn the output off.

Incorrectsensing connection protection (SENSE)

 Turn the output off.

Low AC inputprotection (AC-FAIL)

 Turn the output off.

Shutdown (SD)

 Turn the output off.

Power limit(POWER LIMIT)

 Over power limit.

 Approx. 105% of rated output power

Interface Capabilities

USB

 TypeA: Host, TypeB: Slave, Speed:1.1/2.0, USB Class: CDC(Communications Device Class)

LAN

 MAC Address, DNS IP Address, UserPassword, Gateway IP Address, Instrument IP Address, Subnet Mask

RS-232 /RS-485

 Complies with the EIA232D / EIA485Specifications

GPIB (FactoryOption)

 SCPI - 1993, IEEE 488.2 compliantinterface

Isolated Analog Control Interface (Factory Option)

VoltageControl

 Using 0-5V or 0-10V signals forprogramming and measurement

CurrentControl

 Using 4-20mA current signals forprogramming and measurement

Environmental Conditions

Operatingtemperature

 0˚C to 50˚C

Storagetemperature

 -25˚C to 70˚C

Operatinghumidity

 20% to 85% RH; No condensation

Storagehumidity

 90% RH or less; No condensation

Altitude

 Maximum2000m

Input Characteristics

Nominal inputrating

 100Vac to 240Vac, 50Hz to 60Hz, single phase

Input voltagerange

 85Vac ~ 265Vac

Inputfrequency range

 47Hz ~ 63Hz

Maximum inputcurrent (A)

 21(100Vac) / 11(200Vac)

Inrush current

 Less than50A.

Maximum inputpower (VA)

 2000

Power factor

 0.99 / 0.98

Efficiency(*13) (100Vac / 200Vac) (%)

77 / 79

82 / 85

83 / 86

84 / 87

84 / 87

84 / 87

84 / 87

84 / 87

84 / 87

84 / 87

Hold-up time

 20ms or greater

General

Weight

 Less than 8.7kg (main unit only)

Dimensions

 423(W)×43.6(H)×447.2(D) mm

Cooling

 Forced air cooling by internal fan.

EMC

 Complies with the European EMC directive89/336/EEC for Class A test and measurement products.

Safety

 Complies with the European Low VoltageDirective 73/23/EEC and carries the CE-marking.

Withstand voltage

 AC to Chassis : 1500Vac / 1min

 AC to Output terminal :3000Vac / 1min

 Vout≦150V

 Output terminal to Chassis: 1000Vdc / 1min

 150≦600<>

 Output terminal to Chassis: 1500Vdc / 1min

Insulationresistances

 Chassis and output terminal; chassis andAC input; AC input and output terminal: 100MΩ or more (DC 1000V)