• Key Features:
    -Voltage Range : 150/300/500 Vrms
    -Current Range :
    -Model 66201 - 0.01/0.1/0.4/2 Arms
    -Model 66202 - 0.01/0.1/0.4/2 Arms ; 0.2/2/8/20 Arms
    -Frequency Range : DC, 15Hz~10kHz
    -Embedded high speed DSP, 16 bits Analog/Digital converters
    -10 mA minimum current range & 0.1mW power resolution
    -Meets ENERGY STAR / EN 50564:2011/IEC 62301:2011 / ErP measurement requirements
    -Accumulated energy methods for unstable power measurement
    -User-define criteria provides automatic PASS/FAIL indications
    -Half rack width and 2U height, suitable for system integration
    -Dual current shunt design provides high accuracy over a wide current range (Model 66202)
    -THD and user-specify order distortion measurement (Model 66202)
    -Inrush current and energy measurement (Model 66202)
    -Interface options : USB or USB+GPIB
    -Voltage/ Current harmonics measurement up to 50 orders
  • Key Features:
    -Voltage Ranges:-15/30/60/150/300/600 Vrms/1200Vrms (optional)
    -Current Ranges : 0.005/0.02/0.05/0.2/0.3/0.5/2/5/20/30 Arms
    -Frequency Range : DC, 10Hz~10kHz
    -Embedded high speed DSP, 16 bits Analog/Digital converters withmax samipling rate up to 250kHz
    -Capable of meeting the IEC 61000-4-7 harmonics measurement requirements
    -Smart Range function provides seamless power integration measurement underauto range mode
    -Supports external shunt and CT for higher current applications
    -5 mA minimum current range & 0.1mW power resolution
    -Meets ENERGY STAR/EN 50564 /IEC 62301/ErP requirements
    -User-define criteria provides automatic PASS/FAIL indications
    -THD and user-specify order distortion measurement
    -Inrush current and energy measurements
    -Voltage/ Current harmonics measurements up to 100 orders
    -USB (Host) interface provides data logging functionality
    -Support GPIB, USB, RS232, LAN (option) interface
  • Key Features:
    -Voltage range: 0 ~ 60V
    -Current range: 0 ~ 7A
    -Power range: 0 ~ 150W
    -Wide range of voltage & current combinations in constant power
    -Clean and stable power with programmability at an affordable price
    -Low noise: <3mVp-p
    -High transient response: <30us
    -High-speed programming
    -Precision V&I measurements
    -Standard GPIB/USB interface
    -Remote sense
    -Master-slave parallel and serial and serial control
    -8 steps for auto sequencing programming
    -16 storage locations for user-defined operating states
    -OVP, Current limit, Thermal and Over-power protection
    -CE Certified
  • Key Features:
    -Power range : 7.5kW
    -Voltage range: 0 ~ 30V
    -Current range: 0 ~ 250A
    -Current stability : 2.5mA(10ppm)
    -High-resolution current programming & Measurement: 20bit DAC/24bit ADC
  • Key Features:
    -Voltage range : 0 ~150V/600V/1000V/1500V/1800V
    -AC input voltage range : 200/220Vac, 380/400Vac , 440/480Vac
    -3U/18kW high power density module with easy master/slave parallel operation
    -Fast transient response solar array simulation
    -Simulation of multiple solar cell material's I-V characteristic (fill factor)
    -Simulation of dynamic irradiation intensity and temperature level from clear day to cloud cover conditions
    -Shadowed I-V curve output simulation (up to 4096 data points)
    -Low leakage current (< 3mA)
    -Precision V & I measurements
    -Auto I-V program: 100 I-V curves & Dwell time 1-15,000s
    -Static & dynamic MPPT efficiency test (accumulated energy methods)
    -Data recorded via softpanel
    -Standard USB / RS232 / RS485 interface
    -Optional GPIB / Ethernet interface
    -Real time analysis of PV inverters' MPPT tracking via softpanel
    -Free graphic user interface - softpanel for operation
    -Real world weather simulation fast I-V curve update rate : 1s
    -Support up to ten-channel SAS control for multi-MPPT testing
    -Build-in dynamic MPPT test profile of EN50530, Sandia, CGC/GF004 and NB/T 32004